Evaluation of the Biological Shields of the Secondary Standard Dosimetry Laboratory of Ghana Using MCNP5
P. Deatanyah, G. Emi-Reynolds, J.K. Amoako, C.C. Arwui, S. Wotorchi-Gordon, H. Lawluvi, D.O. Kpeglo and J. Annkah
Corresponding Author: P. Deatanyah
Submitted: 2011 September, 16
Accepted: 2011 October, 15
Published: 2012 March, 15 |
Abstract:
|
The primary objective with radiation sources and facilities is the protection of both radiation workers
and the general public. The biological shields of the Secondary Standard Dosimetry Laboratory of the Radiation
Protection Institute (RPI) Ghana had been evaluated for a collimated isotropic cesium-137 source for calibration
purpose using MCNP5 code. The dose rate at supervised areas ranged from 0.57 to 8.35 :Sv/h and 0.26 to
10.22 :Sv/h at control areas when the source was panoramic. When the source was collimated, the dose rate
ranged from 0.05 to 0.30 :Sv/h at supervised areas and 0.23 to 8.88 :Sv/h at control areas for 22.2 GBq of the
cesium-137 source. The scatter contribution from the surfaces of the walls and roofs were also accounted for.
The scatter radiation in the room decreased to 400 :Sv/h when the source was first collimated and to 3.5 :Sv/h
when the source was further collimated. These results agreed quite well with experimental measurement. To
effectively protect the staff, a narrow beam of 1.2 cm diameter which was defined at 1.0 m by the total surface
of the ISO slab phantom was recommended to reduce the dose rate to less than 1.5 :Sv/h outside the calibration
bunker even when the current activity is doubled. It was concluded that the 4.7 cm diameter of the existing
narrow beam should be decreased to 1.2 cm by further collimation of the beam.
Key words: Ambient dose equivalent, biological shielding, MCNP5, modeling, SSDL, ,
|
Abstract
|
PDF
|
HTML |
|
Cite this Reference:
P. Deatanyah, G. Emi-Reynolds, J.K. Amoako, C.C. Arwui, S. Wotorchi-Gordon, H. Lawluvi, D.O. Kpeglo and J. Annkah, . Evaluation of the Biological Shields of the Secondary Standard Dosimetry Laboratory of Ghana Using MCNP5. Research Journal of Applied Sciences, Engineering and Technology, (06): 544-550.
|
|
|
|
|